Comparative efficacy of terbutaline sulphate delivered by Turbuhaler dry powder inhaler or pressurised metered dose inhaler with Nebuhaler spacer in children during an acute asthmatic episode
  1. S Drblik1,
  2. G Lapierre1,
  3. R Thivierge2,
  4. J Turgeon2,
  5. P Gaudreault2,
  6. B Cummins-McManus2,
  7. I Verdy2,
  8. J Haddon3,
  9. J Lee3,
  10. S Spier4
  1. 1Service of Pulmonology, Hôpital Ste-Justine, Université de Montréal, Québec, Canada
  2. 2Department of Pediatrics, Hôpital Ste-Justine
  3. 3AstraZeneca Canada, Inc.
  4. 4Respiratory Division, Alberta Children’s Hospital, University of Calgary, Calgary, Alberta, Canada.
  1. Correspondence to:
    Dr S Spier, Alberta Children’s Hospital, 1820 Richmond Road, SW Calgary, Alberta, T2T 5C7 Canada;
    sheldon.spier{at}crha-health.ab.ca

Abstract

Aims: To compare the efficacy of terbutaline sulphate delivered via Turbuhaler with a pressurised metered dose inhaler (pMDI) connected to Nebuhaler spacer in a population of asthmatic children presenting to emergency departments because of an acute episode of asthma.

Methods: Randomised double blind, double dummy, parallel study of acute asthma in the emergency department. A total of 112 children (6–16 years), who had a diagnosis of asthma, a baseline FEV1 of 25–60% of predicted normal value (PNV), and the ability to perform spirometry were studied. Patients received two doses of 0.5 mg/10 kg (maximum 2.0 mg) of terbutaline sulphate at time 0 minutes and time 30 minutes. The two groups were also stratified into subgroups based on FEV1: 25–45% and 45.1–60% PNV. FEV1 before treatment and at two 15-minute intervals after each treatment was the main outcome measure. PIF, PEF, heart rate, SpO2, and tremor were also measured at these times.

Results: Both the Turbuhaler and pMDI+Nebuhaler groups showed significant increases from baseline to final value in their FEV1 results, 49% and 50% change from baseline to t = 60 min, respectively (p < 0.001) using last value carried forward. No significant difference was found between the two groups for these results. Subanalysis of the stratified groups revealed similar results. In addition, no significant difference was found in the group and subgroup comparisons for heart rate, SpO2, and tremor.

Conclusion: Results show that Turbuhaler and pMDI+Nebuhaler are similar in terms of benefit and side effects in the treatment of acute moderate to severe asthma attacks in this study population.

  • asthma
  • inhaler
  • nebuhaler
  • spacer
  • turbuhaler
  • DPI, dry powder inhaler
  • FEV1, forced expiratory volume in one second
  • ITT, intention to treat
  • PEF, peak expiratory flow
  • PIF, peak inspiratory flow
  • pMDI, pressurised metered dose inhaler
  • PNV, predicted normal value
  • SpO2, transcutaneous oxygen saturation
  • t, time

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

The conventional first line treatment of acute asthma for the paediatric population is a β2 agonist administered in an aqueous solution by nebulisation. In recent years, β2 agonists given by pressurised metered dose inhalers plus a spacer have been shown to be as efficacious as when given via a nebuliser to children with acute asthma.1–5 These findings suggested that other treatment modalities could offer the same advantages and be used as alternatives in the emergency setting in treating patients with acute asthma.

Dry powder formulations of β2 agonists are a well established treatment of children with clinically stable asthma and have been shown to be as efficacious as the pressurised metered dose inhaler (pMDI) with a spacer.6–9 Only recently, the use of dry powder inhalers has been investigated as a possible alternative for acute asthma; however, these studies generally have been in adults.10,11 To our knowledge, only two investigations to date have attempted to address this question in the paediatric population using a limited number of subjects.12,13 The reluctance to use an effort dependent breath actuated device in an acute setting may be due, in part, to the belief that either inspiratory flow is sufficiently compromised during an acute exacerbation or children may not be able to perform the inspiratory manoeuvre correctly, thereby affecting the amount of medication being effectively delivered.

We designed this study, therefore, to compare one such dry powder inhaler (DPI), Turbuhaler (AstraZeneca), with a pMDI and a spacer, Nebuhaler (AstraZeneca), in their efficacy of delivering equally effective doses of terbutaline sulphate to children presenting to the emergency department for first line treatment of an acute asthmatic episode.

METHODOLOGY

Patients

Patients aged 6–16 years inclusive, with acute asthma, who presented to one of three emergency units in Canada, were recruited to the study: Hôpital Ste-Justine, Montréal, Québec (n = 102); Alberta Children’s Hospital, Calgary, Alberta (n = 7); and Misericordia Community Hospital, Edmonton, Alberta (n = 3). Inclusion criteria were a baseline forced expiratory volume in one second (FEV1) of 25–60% of predicted normal value (PNV),14 baseline peak inspiratory flow (PIF) of at least 30 l/min when measured through Turbuhaler, and the ability to perform correctly both spirometry and the inhalational techniques required of Turbuhaler and pMDI+Nebuhaler. In addition, baseline resting heart rate could not exceed 160 beats/minute.

Patients were ineligible if they had used inhaled β2 agonists within three hours prior to enrolment, or theophylline, ipratropium bromide, or oral β2 agonists within 12 hours. Potential patients were also excluded if they had any known contraindication to sympathomimetic amines, lactose (non-drug component of the placebo), or pMDI additives (Freon 11, 12, 114, and sorbitan trioleate).

The respective ethics committee of each participating hospital approved the study, and informed written consent was obtained from the parents. The study was performed in accordance with the principles stated in the Guidelines on Research Involving Human Patients issued by the Medical Research Council of Canada, which encompass the Declaration of Helsinki of 1975 (revised 1983), and with Good Clinical Practice.

Study design

A double blind, double dummy, randomised, parallel design was used. Patients were first stratified into two groups based on their initial FEV1 expressed as the percent PNV: 25–45% and 45.1–60%. In addition, within each stratified group the treatment and delivery system were randomised and were alternated to avoid a potential bias based on the order of treatment and device.

Patients were randomised into two groups to receive either 0.5 mg/10 kg (maximum 2.0 mg) terbutaline sulphate via Turbuhaler and placebo pMDI+Nebuhaler spacer (or the reverse order) or 0.5 mg/10 kg (maximum 2.0 mg) terbutaline sulphate via the pMDI+Nebuhaler spacer and placebo Turbuhaler (or the reverse order). The treatment was given twice within the 60 minute treatment period, once at the beginning of the study, t = 0, and once again after 30 minutes, t = 30. Approximately 10 minutes prior to commencing the study, each patient received a dose of oral prednisone. Patients weighing <25.0 kg received 25 mg prednisone, while patients weighing ⩾25.0 kg received 50 mg.

All patients were shown the required inhalation techniques for each device and were required to show the techniques prior to commencing the study. For Turbuhaler, patients were instructed to inhale as fast and as deeply as possible from functional residual capacity followed by a natural breath hold lasting approximately three seconds.15 For the pMDI+Nebuhaler spacer, five breaths of sufficient force to move the valve in Nebuhaler were required for each puff from the pMDI.16 All treatments were given with the patients in the sitting position.

Measurements

Spirometry was performed on a Vitalograph Compact II to measure the variables peak inspiratory flow (PIF) and FEV1. Three technically correct forced expiratory manoeuvres were executed from which the best result was chosen, in accordance with all but one of the American Thoracic Society guidelines.17 A difference of 8% instead of 5% in FEV1 among the three test results was accepted since the acute state of the patient would not permit excessive manoeuvres for fear of risking bronchospasm or attrition.

Measurements of PIF, both with and without Turbuhaler, were also made in triplicate using the Micro Plus Spirometer modified for inspiratory flow. Peak inspiratory flow through Turbuhaler was obtained by attaching Turbuhaler in series to the spirometer by a specially designed adapter. Recordings of PIF through Turbuhaler were obtained during treatment delivery at t = 0 minutes and t = 30 minutes. All other spirometry assessments were completed at baseline and following both treatments at 15 minute intervals, which resulted in recordings at time 0, 15, 30, 45, and 60 minutes.

Heart rate and SpO2 were measured using a pulse oximeter. Hand tremor was assessed using predefined criteria and recorded as absent, mild, moderate, or severe. The data were collected from each patient prior to spirometry, with the patient having assumed a semirecumbent position for at least five minutes.

The right to terminate the study before completion was an option that the patient, parent, or physician could exercise at any time. In addition, patients were withdrawn from the study if (1) their baseline FEV1 did not increase by more than 10% 30 minutes after the first treatment; (2) their baseline heart rate increased by more than 40%; or (3) their tremor was assessed as severe.

Statistical analysis

The data were analysed with SAS V6.12/Windows NT. The change from baseline to the last FEV1, PIF with and without Turbuhaler, PEF, and safety variables (heart rate, tremor, and SpO2) were compared between the treatment groups by an analysis of covariance (ANCOVA) with factors for treatment, stratification, and interaction effects between treatment and stratification in the preliminary model. The corresponding baseline value was used as the covariant. Where spirometry data were missing, the principle of last value carried forward was used. The interaction effect was removed from the final model because of statistical insignificance. The 95% confidence intervals for treatment differences were calculated. No analysis for centre effect was undertaken because of the small sample sizes at two of the three centres.

The sample size calculation was based on comparison of the mean change in FEV1 %PNV from baseline between treatments. With a two sided test at α = 0.05 and a power of 80%, 64 patients were required in each group in order to detect a clinically relevant difference of 5% in the FEV1 with a standard deviation (SD) of 10%. Since the study was terminated prematurely because of a slow recruitment rate, a retrospective sample size calculation was performed. The actual power of the study was 75% given a total of 112 patients and SD of 10%. All randomised patients were included in the intention to treat analysis (ITT). All statistical tests were two sided and p values of <0.05 were considered statistically significant.

RESULTS

Of the 112 patients that were enrolled into the study, nine children did not complete the investigation. Six patients were from the Turbuhaler group, of which three were unwilling to continue the study and three were removed after 30 minutes when their baseline FEV1 failed to increase by more than 10%. Their ages ranged from 6 to 10 years, and their baseline FEV1 values were 36–60% PNV. The reasons for early withdrawal for the three patients in the pMDI+Nebuhaler group were unwillingness to continue in two patients and a <10% increase in FEV1 after 30 minutes following the first treatment in one patient. These three patients were between 10 and 13 years of age with baseline FEV1 values 28–53% PNV. Since the ITT approach was selected for analysing the data, all data were included in the analysis.

Table 1 shows the demographic and baseline spirometric data of the 112 patients who were enrolled in the study. No significant difference was found between the two groups for any baseline variable, indicating that both groups were well matched for sex, age, height, and FEV1. The mean baseline FEV1 (SD) value for the Turbuhaler group was 44.8% (9.0%) PNV, and the mean baseline FEV1 (SD) value for the pMDI+Nebuhaler group was 43.4% (10.1%) PNV. Stratification of the two treatment arms into the two subgroups (FEV1 25–45% and 45.1–60%) resulted in comparable findings. In the FEV1 25–45% stratum, there were 26 patients in the Turbuhaler group (FEV1 %PNV: 36.6 (4.9); range 26–44) and 26 patients in the pMDI+Nebuhaler group (FEV1 %PNV: 34.1 (5.1); range 26–45). In the FEV1 45.1–60% stratum, there were 31 patients in the Turbuhaler group (FEV1 %PNV: 51.6 (5.0); range 43–60) and 29 in the pMDI+Nebuhaler group (FEV1 %PNV: 51.8 (4.4); range 46–60). Neither these differences nor any other baseline measurements resulting from the stratification proved to be significantly different between the two groups.

Table 1

Demography of patient population

Similar doses of terbutaline sulphate were given to both groups at the two dosing times. At the initial treatment time (t = 0), the Turbuhaler group received a mean value of 1.75 (0.33) mg (range 1.0–2.0 mg), while the pMDI+Nebuhaler group received a mean value of 1.71 (0.3) mg (range 1.0–2.0 mg) of terbutaline sulphate. At the second treatment time (t = 30), the Turbuhaler group received an additional mean amount of 1.78 (0.32) mg (range 1.0–2.0 mg), and the pMDI+Nebuhaler group received a mean amount of 1.69 (0.33) mg (range 1.0–2.0 mg). None of these differences were significantly different between the two groups.

Comparison between the two groups showed no significant difference for the changes in FEV1, the primary variable, from baseline to final value (p > 0.05; 95% CI −4.4% to 3.5% PNV). Each group did show, however, a significant improvement after each treatment when compared with baseline values (p < 0.001). As fig 1 shows, 30 minutes following the first treatment dose at t = 0, there was a 33% increase in the FEV1 %PNV in the Turbuhaler group and a 35% increase in the pMDI+Nebuhaler group. After the second and final treatment dose given at t = 30, additional increases of 21% and 16% were found in the Turbuhaler and Nebuhaler groups respectively, resulting in an overall change from baseline to final value of approximately 49% for the Turbuhaler group and 50% for the Nebuhaler group.

Figure 1

FEV1 % PNV over time (mean ± 2SEM); p < 0.05 change from baseline to last value within each group. The differences between groups were not significant.

Figure 2 represents the mean FEV1% PNV for the two groups stratified into their two subgroups according to baseline FEV1% PNV: 25–45% and 45.1–60%. There was no significant difference in baseline values between the two groups within the 25–45% stratum, and the final mean value after the two treatment doses at t = 60 was 62.1% (12.3%) PNV for the Turbuhaler group and 55.7% (13.0%) PNV for the pMDI+Nebuhaler group. This results in a mean absolute change of 26% and 22%, respectively. While these increases were statistically significant within each treatment stratum, no difference was observed between the Turbuhaler and the pMDI+Nebuhaler within the 25–45% stratum. Similar statistical results were found for the 45.1–60% stratum. The final mean value at t = 60 minutes was 70.8% (8.1%) PNV for the Turbuhaler group and 73.1% (8.8%) PNV for the pMDI+Nebuhaler group, which yielded a mean absolute change of 19% and 21%, respectively. Once again, although these increases proved to be statistically significant within each treatment subgroup, no difference was found between the Turbuhaler and pMDI+Nebuhaler 45.1–60% stratum. Similar statistical results were found for PIF and PEF.

Figure 2

FEV1 % PNV by stratum over time (mean ± 2SEM).

Figure 3 shows the relation between the bronchodilatory effects of the terbutaline sulphate delivered in dry powder and the inspiratory flow through Turbuhaler. Only the data from the Turbuhaler group are shown. No significant correlation could be found between the percent change from baseline to the final FEV1 value and the PIF measured through Turbuhaler during the first treatment dose at t = 0 (p = 0.10; r = 0.31). Similar results were found when this relation was examined using the PIF measured through Turbuhaler during the second treatment dose at t = 30. The six data points below the 10% mark on the y axis are the results of the six patients who were enrolled and randomised into the Turbuhaler group but did not complete the protocol.

Figure 3

Percent change in FEV1 versus baseline PIF via Turbuhaler device for the Turbohaler group.

The changes for heart rate from baseline to final value were statistically significant (p < 0.001) within each group, but no significant difference was found when between group comparisons were made (table 2). The greatest increase in heart rate for both groups occurred 30 minutes after the first treatment (Turbuhaler: 12% change; pMDI+Nebuhaler: 10% change) which was followed by another increase, albeit smaller, 30 minutes after the second treatment (Turbuhaler and pMDI+Nebuhaler: 7% change). No patient in either group had an increase in heart rate of more than 40% above their baseline value. The changes in SpO2 from baseline to final value were not significantly different between the two groups, and although within group comparisons were statistically significant, there was no clinical relevance to these changes.

Table 2

Oxygen saturation and heart rate data shown by group and time post-treatment

On entry into the study, 88% of the Turbuhaler patients and 91% of the pMDI+Nebuhaler patients did not exhibit any hand tremor. At the final measurement interval of the study, after two treatment doses had been given, mild hand tremor was observed in 26% and 56% of the Turbuhaler and pMDI+Nebuhaler patients, respectively, while 61% of the Turbuhaler and 40% of the pMDI+Nebuhaler patients were exhibiting moderate hand tremor. Statistical analysis of these differences did not show any significance for the between group comparisons, while within group comparisons did reach significance (p < 0.001). At no time during the study did any patient exhibit a severe hand tremor.

DISCUSSION

The results from this study have shown that in our sample population terbutaline sulphate delivered in dry powder by Turbuhaler was as efficacious as the same nominal dose of terbutaline sulphate delivered by the pMDI+Nebuhaler spacer as a first line treatment of moderate and severe acute asthma attacks in children. These findings are in agreement with previous reports that suggested that Turbuhaler was an effective method of treating children during an acute asthma exacerbation.12,13 Springer and colleagues13 compared 2.5 mg terbutaline sulphate delivered via Turbuhaler with 5.0 mg terbutaline sulphate given by nebulisation and found significant and comparable improvements in PEF in a sample of 12 children (aged 7–12 years) attending an emergency department for treatment of an acute asthma exacerbation. In the other paediatric study, Rufin and colleagues12 recruited 30 children (aged 6–14.5 years) with moderate acute asthma from an outpatient clinic. Using an open design, two groups of 15 patients received a single dose of 0.5 mg terbutaline sulphate either via Turbuhaler or by a pMDI+Nebuhaler. This study also found significant and comparable improvements in FEV1 between the two groups up to 30 minutes following the treatment.

While the conclusions of these studies are in agreement with our study, neither study reported their statistical power. Therefore, the possibility exists that their sample sizes were too small to detect a significant treatment difference. Although the actual power of the current study was calculated to be 75% instead of the initially planned 80%, this was not a concern since the confidence interval of the treatment difference had a small range (that is, both the lower and upper limits were less than 5%). If we had chosen a value of 7% instead of 5% for the clinically relevant difference in FEV1 as significant, the power of our study would increase to 96%. In fact, a clinically relevant difference as high as 12% in the FEV1 PNV has been used previously in a comparably designed emergency department study which examined the difference between β2 agonists delivered by nebuliser and pMDI with a spacer in children with mild to severe acute asthma.5 In addition, we chose to compare the Turbuhaler with a metered dose inhaler and spacer because of the simplicity and ease of use of the spacer, as well as results of numerous studies, including that of Kerem and colleagues,5 showing the equivalency of the spacer to the wet nebuliser.

The lack of correlation between the percent increase in FEV1 from baseline to final value and the measured inspiratory flow through Turbuhaler (fig 3) suggests that the bronchodilatory effects achieved with the DPI were independent of inspiratory flow above our minimum requirement of 30 l/min. Similar observations were reported by Pedersen and colleagues.18 Their investigation examined the bronchodilatory effects of a submaximal dose of terbutaline sulphate delivered via Turbuhaler while varying inspiratory flows between 13 and 60 l/min in a sample of 14 children (aged 7–15 years) with stable chronic asthma. They found the increase in FEV1 after treatment was comparable for patients with mean inspiratory flows of 31–60 l/min. However, when the inspiratory flow was decreased to 22 l/min., the treatment effect was reduced by approximately 67%. In addition, a further decrease in inspiratory flow to 13 l/min resulted in a 33% treatment effect. Based on their findings, we chose a minimum inspiratory flow through Turbuhaler of 30 l/min as an inclusion criterion, and it would appear that our results in an unstable acute asthmatic state are in agreement with their results in a stable chronic asthmatic state.

In our study, we did not find inspiratory flow to be a limiting factor in either patient attrition or recruitment. Once enrolled, no patient was withdrawn from the study because of their inability to continue to generate throughout the course of the protocol the minimally required inspiratory flow through Turbuhaler. This was observed in children as young as 6 years and in children with a baseline FEV1 as low as 25%. In addition, apart from the 112 patients enrolled, only five patients (4%), aged 6–13 years, whose baseline FEV1% PNV was 25–60%, were not enrolled when they failed to achieve a Turbuhaler PIF of at least 30 l/min; values were 24–28 l/min. In this very small group, this was a result of either limited ability to perform the technique or actual flow limitation, and does not necessarily mean they would not have improved using this technique. This small number of subjects not studied should not detract from the major findings of this study. Of interest, studies investigating Turbuhaler in adults also found that only 2–4% of their sample were unable to generate 30 l/min, and none of them were lower than 25 l/min.11,19 Although these patients were not enrolled in either their studies or in our study, it is quite probable that a beneficial response from the treatment would have been observed based on the findings from Pedersen and colleagues18 which showed that a significant bronchodilatation could still be achieved with Turbuhaler inspiratory flows of 22 l/min.

The similar bronchodilatory improvements accounted for by the changes in FEV1 not only were observed between the two groups, but also were comparable within the strata subgroups. The results of stratification of the sample into severe (25–45% FEV1 PNV) and moderate (45.1–60% FEV1 PNV) indicated that the bronchodilatory benefits were not related to degree of severity. Indeed, the mean absolute changes within each stratum were similar. However, when calculated on a relative basis, improvements in the severe stratum were greater than those changes calculated for the moderate stratum. One possible explanation for this observation could be that the severe patients had a greater potential to improve based on their lower initial value. The profile of our sample population could also account for the variation in bronchodilatory response. The duration of the asthmatic episode previously prescribed, type and dose of medication, and effective treatment delivery modalities were uncontrolled factors, which could have accounted for the degree of response to our treatment. Our patients were selected essentially on degree of severity, capabilities, and previous medication restrictions.

The calculated dose of terbutaline sulphate used for each patient was the same for the two delivery methods, and based on the FEV1 results, provided similar bronchodilatory effects. On this topic the literature is equivocal. Previous cumulative dose response adult20 and children6,9 studies, as well as investigations comparing predetermined equivalent doses at various time intervals7,8,21 have found no significant difference in efficacy between Turbuhaler and pMDI. On the other hand, other well designed studies have shown a higher dose-potency relation for Turbuhaler.10,22–24 Moreover, Nana et al found that in their adult emergency unit study there was equal efficacy between the Turbuhaler and pMDI with Volumatic spacer, even though the pMDI contained twice the dose of salbutamol as did the Turbuhaler.24 The lack of agreement among the studies may be explained by their methodological differences. In our study, it was not our intent to evaluate a dose response to terbutaline sulphate but to initiate an effective treatment for an acute asthmatic episode based on our emergency department procedure which is to give a high dose of a β2 agonist. The possibility exists therefore that our initial dose brought our patients nearer the flatter portion of the dose-response curve where differences between methods would not be detected.

Finally, our findings showed that the treatment benefits to side effects ratio were comparable between the two groups. The equal bronchodilatation, as indicated by the FEV1 results, achieved with both devices was accompanied by similar increases in heart rate and tremor within both groups. In addition to the finding that no patient showed either a 40% increase in baseline heart rate or a severe tremor during the treatment protocol, no patient requested to be withdrawn from the study because of an intolerance of these effects.

Conclusion

Terbutaline sulphate delivered in dry powder formulation via Turbuhaler was as efficacious as the pMDI+Nebuhaler spacer in the treatment of moderate to severe acute asthma attacks in our sample of children presenting to the emergency department. This observation was seen in children as young as 6 years of age, as well as in patients with an FEV1 percent PNV as low as 25%, both factors not affecting their ability to perform correctly the inspiratory manoeuvre required of Turbuhaler. Our findings suggest that the Turbuhaler (DPI) is an effective alternative modality for the first line treatment of an acute asthmatic episode in children.

Acknowledgments

We wish to express our gratitude to Drs Benoit Bailey, Sylvie Bergeron, Perrette Payée, David Schragge, and Cheri Nijssen-Jordan for their assistance in the recruitment phase of the study. We would also like to acknowledge the support of AstraZeneca Canada, Inc.; Mr Michael Bukoski for help in preparing the protocol; and Dr Robert Jenkins for assistance in reviewing and preparing the manuscript.