Stratus CS cardiac troponin I method: performance characteristics including imprecision at low concentrations

Clin Biochem. 2004 Aug;37(8):679-83. doi: 10.1016/j.clinbiochem.2004.03.009.

Abstract

Objectives: Define performance characteristics of the Stratus CS cardiac troponin I (cTnI) assay in routine practice.

Methods: Imprecision at low cTnI values was assessed according to NCCLS EP5-A. Linearity was assessed by regression analysis; the reference interval was determined in 345 healthy individuals. Minimum detectable concentration (MDC) was assessed with 20 replicates of zero calibrator, and method correlation and bias (n = 77) were by Passing-Bablok regression and Bland-Altman plot.

Results: cTnI of 0.06 microg/l corresponded to the 10% CV; the 99th percentile of the reference population was 0.07 microg/l. Linearity extended from 0 to 50.0 microg/l and the MDC was 0.010 microg/l. The method comparison equation was Stratus CS = 1.00 (Dimension RxL)-0.02 microg/l with no bias.

Conclusions: The Stratus CS cTnI assay in routine practice has performance characteristics appropriate for clinical use, including good correlation to a central laboratory cTnI method and imprecision of a high sensitivity troponin method with a CV of <10% at the 99th percentile of the reference population.

Publication types

  • Comparative Study

MeSH terms

  • Calibration
  • Chemistry, Clinical / instrumentation*
  • Chemistry, Clinical / methods*
  • Humans
  • Myocardium / metabolism*
  • Reference Values
  • Regression Analysis
  • Troponin I / analysis*