Fluoroscopically Guided Implantation of Modern Cardiac Resynchronization Devices: Radiation Burden to the Patient and Associated Risks

https://doi.org/10.1016/j.jacc.2005.01.070Get rights and content
Under an Elsevier user license
open archive

Objectives

To establish radiation risks for patients undergoing fluoroscopically guided cardiac resynchronization device implantation.

Background

Cardiac resynchonization therapy (CRT) may be associated with extended fluoroscopic exposure.

Methods

The fluoroscopy time, dose-area product (DAP), exposure parameters, and percentage contribution of the fluoroscopic projections commonly used were recorded in a series of 14 consecutive patients referred for cardiac resynchronization device implantation and compared to corresponding data obtained from a control group of 20 patients who underwent a conventional rhythm device implantation operation. The DAP to peak skin dose, DAP to effective dose, and DAP to gonadal dose conversion factors were determined for biventricular pacing and conventional rhythm device implantation using a humanoid phantom and thermoluminescence dosimetry.

Results

The mean total fluoroscopy time and DAP values were 35.2 min and 4,765 cGy cm2, respectively, for biventricular pacing and 8.2 min and 1,106 cGy cm2, respectively, for conventional rhythm device implantation. Patient skin dose from biventricular pacing procedures requiring extended fluoroscopic exposure may exceed threshold dose for the induction of skin effects only if X-ray source-to-skin distance is kept low. The risk values for fatal cancer and severe hereditary disorders, respectively, associated with a typical CRT procedure were 273 per million and 0.2 per million treated patients.

Conclusions

Radiation risks associated with fluoroscopically guided CRT procedures may be considerable. Present data may be used for the estimation of patient radiation risks from CRT procedures performed in other institutions.

Abbreviations and Acronyms

cGy cm2
centigray × centimeter2
CRT
cardiac resynchronization therapy
DAP
dose-area product
ICD
implantable cardioverter-defibrillator
ICRP
International Commission on Radiological Protection
LAO
left-anterior oblique
PA
posteroanterior
RAO
right-anterior oblique
SSD
source-to-skin distance
TLD
thermoluminescence dosimeter
Sv−1
sievert−1
Gy−1
gray−1

Cited by (0)