Table 3:

Results of the meta-analysis of studies including the primary parameters of true-positive, false-positive, true-negative and false-negative

Location, apnea–hypopnea cut-offNo. of studiesOverall heterogeneitySensitivity (95% CI)Specificity (95% CI)Area under the ROC curve (95% CI)Positive LR (95% CI)Negative LR (95% CI)
I2 (95% CI)p value for Q statistic
Home, ≥ 5 events/h846,47,52,55,84,85,89,10200.50.93 (0.90–0.95)0.60 (0.51–0.68)0.89 (0.86–0.92)2.3 (1.9–2.9)0.11 (0.07–0.16)
Laboratory, ≥ 5 events/h747,52,61,65,67,70,11885 (68–100)0.0010.96 (0.90–0.98)0.76 (0.63–0.85)0.92 (0.90–0.94)3.9 (2.6–6.1)0.05 (0.02–0.13)
Home, ≥ 10 events/h646,47,55,76,89,9153 (0–100)0.060.83 (0.73–0.89)0.81 (0.70–0.89)0.89 (0.86–0.91)4.3 (2.7–7.0)0.22 (0.14–0.33)
Laboratory, ≥ 10 events/h647,48,58,61,65,7000.30.92 (0.87–0.95)0.85 (0.77–0.90)0.93 (0.91–0.95)6.0 (4.0–8.9)0.09 (0.05–0.15)
Home, ≥ 15 events/h84648,52,55,85,89,10482 (62–100)0.0020.79 (0.71–0.86)0.79 (0.63–0.89)0.85 (0.82–0.88)3.7 (2.1–6.7)0.26 (0.19–0.37)
Laborator, ≥15 events/h947,48,52,56,58,61,65,67,7066 (23–100)0.030.92 (0.86–0.96)0.91 (0.85–0.95)0.97 (0.95–0.98)10.6 (6.1–18.2)0.08 (0.04–0.15)
Home, ≥ 30 events/h548,52,55,88,10400.40.79 (0.72–0.85)0.90 (0.84–0.95)0.86 (0.83–0.89)8.2 (4.7–14.6)0.23 (0.16–0.32)
Laboratory, ≥ 30 events/h448,52,58,6700.50.97 (0.92–0.99)0.93 (0.89–0.96)0.99 (0.98–1.00)14.9 (8.6–25.8)0.03 (0.01–0.08)
  • Note: CI = confidence interval, LR = likelihood ratio, ROC = receiver operator characteristic.